Site Preparations
2-18 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Environmental Dangers
Patient Vicinity UL60601-1 (USA)
Commercial devices such as laser cameras, printers, VCRs and external monitors,
usually exceed allowable leakage current limits and, when plugged into separate AC
outlets, are in violation of patient safety standards. Suitable electrical isolation of such
external AC outlets, or providing the device with extra protective earth, will be required
in order to meet UL60601-1 and IEC60601-1 / IEC60601-1-1 standards for electrical
leakage.
2.12.20DV (UL60601-1:2003)
In area in which patients are normally cared for, the patient vicinity is the space with
surfaces likely to be contacted by the patient or attendant who can touch the patient.
This encloses a space within the room 1.83 m (6 ft.) beyond the perimeter of the bed
(examination table, dental chair, treatment booth, and the like) in its intended location,
and extending vertically 2.29 m (7.5 ft.) above the floor.
1. Patient environment

Table of Contents

Other manuals for GE Vivid E90

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E90 and is the answer not in the manual?

GE Vivid E90 Specifications

General IconGeneral
BrandGE
ModelVivid E90
CategoryMedical Equipment
LanguageEnglish

Related product manuals