Lockout/Tagout (LOTO) requirements
Vivid E80/E90/E95 – Service Manual 1-17
GC091052 Rev. 3
Lockout/Tagout (LOTO)
requirements
Follow Lockout/Tagout requirements by ensuring you are in total
control of the AC power plug at all times during the service
process.
To apply Lockout/Tagout (LOTO):
1. Plan and prepare for shutdown.
2. Shutdown the equipment.
3. Isolate the equipment.
4. Apply Lockout/Tagout Devices.
5. Control all stored and residual energy.
6. Verify isolation.
All potentially hazardous stored or residual energy is relieved.
WARNING
Energy Control and Power Lockout for Vivid E80/E90/E95.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.

Table of Contents

Other manuals for GE Vivid E90

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E90 and is the answer not in the manual?

GE Vivid E90 Specifications

General IconGeneral
BrandGE
ModelVivid E90
CategoryMedical Equipment
LanguageEnglish

Related product manuals