151
Using Instrument and Disk Storage
Copying Configuration from Disk to Instrument Store
Copying Configuration from Disk to
Instrument Store
Description If you have a configuration stored in the instrument store that you wish
to use on another instrument, you can copy it to Disk. The configuration
can then be downloaded from the disk in another HP 37717C with the
same options as the original instrument.
HOW TO: 1 Choose , DISK OPERATION [FILE] [COPY]
[CONFIGURATION] and enter the Instrument Store number.
If you wish you can enter a description of the configuration. See
"Managing Files and Directories on Disk " page 140.
The description can contain up to 24 alphanumeric characters.
2 Enter the filename the configuration is to be copied from. See
"Managing Files and Directories on Disk " page 140.
The file name can contain up to 8 alphanumeric characters.
The filename extension is fixed as .CNF.
3 To copy the configuration from Disk to instrument, choose .
If you have entered a instrument store number which already contains
a configuration, a warning "Are you sure you wish to continue" is
displayed.
If YES is selected, the data in the instrument store will be overwritten.
If NO is selected, the operation is aborted.
OTHER
FLOPPY DISK
OK
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for HP 37717C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 37717C and is the answer not in the manual?

HP 37717C Specifications

General IconGeneral
BrandHP
Model37717C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals