I
-
Special Functions
2
-
CW PEAKNG
2-cw
This function causes a one-time execution of the peaking function when the
PEAKNG
sweeper is in CW or manual sweep mode. It aligns the output
lilter
so that its
passband is centered on the RF output.
CW peaking is used to obtain the maximum available power and spectral
purity, and the best pulse and FM envelopes at a given frequency.
NOTE
If the instrument is not in CW or manual sweep mode, nothing will happen when
(Hz/s/ENTER_)
is
pressed.
Associated SCPI Commands
CALibration:PEAKing[:EXECute]
Executes the peaking function.