Error Messages
SCPI Error Messages
606
“Warning! Learn String FW Revision Not
Matched(606)”
The learn string that was received does not match the current
firmware revision. It was rejected and not used.
607
-200, “Execution error;Execution Not Allowed. Currently In
Restricted
Mode(607)”
The instrument is in a restricted mode due to either the operation
of self test, or current operation of a calibration. Running of most
commands is not allowed. For best results, a device clear followed
by a *rst command should be sent.
608
-200, “Execution error;Execution Not Allowed. Currently In
Network Analyzer
mode(608)”
Frequency zero can only be executed when the instrument is
in stand alone mode. When the instrument is connected to an
IIF’
8757 or when the instrument is speaking 8350 compatibility
language, frequency zero cannot be implemented. If the user
attempts to implement frequency zero in these modes, an error
message will be generated.
Calibration
Routine
Errors
701
702
703
-300, “Device specific error;Peaking
Failed(701)”
For unspecified reasons, the CW peaking algorithms failed.
-300, “Device specific error;Peaking Never
Leveled(702)”
ALC could not achieved level power under the current conditions.
-300, “Device specific
error;Instrument
Not in CW
Mode(703)”
It is required that the instrument first be in CW mode before a YTF
peak is executed.
16-33