EasyManuals Logo
Home>HP>Blower>83751A

HP 83751A User Manual

HP 83751A
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #233 background imageLoading...
Page #233 background image
I
-
I
-
Error Messages
SCPI Error Messages
714
-300, “Calibration Array Elements Sent In Descending
Order(714)”
A calibration correction flatness array was sent in descending
order. The new array is rejected at the point of the descending
element, X, which causes the array to be out of order. The
previous elements, up to element X, are being written over by the
new elements and cannot be restored.
715
-300, “Entered Password does not match the Security
Password(715)”
The user is trying to change the calibration security password and
the verified password is incorrect as it does not match the system
security password.
716
-300, “User-Defined Password must be a
5numericaldigit(716)”
The user is trying to change the calibration security password and
the new password is not a
5digit
numerical entry.
Loops
Unlocked
Errors
801
802
803
804
-300, “Device specific error;YIG Oscillator
Unlock(801)”
Phase lock with the YIG oscillator was lost or could not be
achieved.
-300,
“Device specific
erroqlleference
Oscillator
Unlock(802)”
Phase lock with the Reference oscillator was lost or could not be
achieved.
-300,
“Device specific error;Fractional-N VCO
Unlock(803)”
Phase lock with the Fractional-N VCO was lost or could not be
achieved.
-300,
“Device specific error;Heterodyne Oscillator
Unlock(804)”
Phase lock with the Heterodyne oscillator was lost or could not be
achieved.