Error Messages
SCPI Error Messages
Bus
Control
Errors
201
204
205
206
207
208
-310, “System
error;Another
Controller Is On The
HP-II3
Bus(201)”
Indicates that during a Flatness Calibration, the instrument was
trying to establish the control of the Power Meter, but Egured out
another controller is on the
HP-E3
bus. Flatness Calibration is
aborted.
-310, “System error;Command Send Error-No HP-IB Devices
Found(204)”
Indicates that during a Flatness Calibration, the instrument was
sending a command to an HP-IB device, but could not End it.
Flatness Calibration is aborted.
-310, “System
error-Cannot
Find Power Meter On HP-IB
Bus(205)”
Indicates that during a Flatness Calibration, the instrument was
trying to establish the control of a supported Power Meter, but
could not End it. Flatness Calibration is aborted.
-310, “System error;Meter Returns Error
Message(206)”
Indicates that during a Flatness Calibration, the HP-IB Power Meter
error checking returns an error message of some type.
-310, “System error;Meter Data Measured Is Invalid or Out Of
Range(207)”
Indicates that during a Flatness Calibration, a reading return value
which HP-IB Power Meter measured is invalid or out of range.
Flatness Calibration is aborted.
-310, “System error;Unable To Receive Message From
Meter(208)”
Indicates that during a Flatness Calibration, a time out is happened
while the instrument was waiting to receive a message from the
Power Meter. Flatness Calibration is aborted.