EasyManuals Logo
Home>HP>Measuring Instruments>8559A

HP 8559A User Manual

HP 8559A
441 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #183 background imageLoading...
Page #183 background image
Page
1
-
3:
Table
1-1.
HP
85596
Specifications
(1
of
4)
2236A
&
Below Change
"
Residual FMw specification to read as follows:
less than 1 kHz p
-
p for a time interval less than or equal to
.O1
sec, 100/120 line voltages; less than 2 kHz p
-
p, 220/240
line voltages.
Delete the following under
"
Maximum Input (without damage)
Levelstt
:
Peak Pulse Power
+50
dBm (100W, 10 microsecond pulse width, 0.01% duty cycle)
with input attenuation
>=30 dB.
Change
"Gain Compression
w
specification to read as follows:
Gain compression
is
less than 0.5 dB for a 0 dBm input level
with
0
dB input attenuation.
Under
*'Display Fidelity
w
, change the Linear specification to read
as follows:
<+
-
0.1 division over full 8 division deflection.
Change
"
Humidity Range (Operating)
"
to read as follows:
<95%
R.H.
0
-
degrees
C
to +40-degrees
C.
Change
"EMIft
to read as follows:
Conducted and radiated interference
is
within the requirements
of methods
CEO3 and RE02 of MIL STD 461A,
VDE
0871 and CISPR
Publications
1,
2, and 4.
2320A
&
Below Change
"
Residual FMw specification to read as follows:
less than 1 kHz p
-
p in 0.1 second.

Table of Contents