2600AS-901-01 Rev. B / September 2008 Return to Section Topics 10-29
Series 2600A System SourceMeter® Instruments Reference Manual Section 10: Triggering
Rising edge master trigger mode
Use the rising edge master (RisingM) trigger mode (see Figure 10-13) to synchronize with non-
Keithley instruments that require a high pulse. Input trigger detection is not available in this trigger
mode. You can use the RisingM trigger mode to generate rising edge pulses.
NOTE The RisingM trigger mode does not function properly if the line is
driven low by an external drive.
Figure 10-13
RisingM output trigger
Output characteristics:
• Configured trigger events as well as the digio.trigger[N].assert and
tsplink.trigger[N].assert commands cause the physical line state to float high
during the trigger pulse duration.
• An action overrun occurs if the physical line state is high while a source event occurs.