2600AS-901-01 Rev. B / September 2008 Return to Section Topics 19-101
Series 2600A System SourceMeter® Instruments Reference Manual Section 19: Remote Commands
setup
The functions and attribute in this group are used to save/recall setups and to set the power-on setup.
setup.poweron
Attribute The saved setup to recall when the unit is turned on.
Default 0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage n = setup.poweron -- Reads the power-on setup.
setup.poweron = n -- Writes the power-on setup.
n Setup number to recall on power up (0 to 5).
Remarks •For an n setting of 0, the unit powers up to the factory default (reset) setup. For an
n setting of 1 to 5, the unit powers up to a user saved setup.
Details See Power-up in Section 1.
Example Sets unit to power on to the factory defaults:
setup.poweron = 0
setup.recall
Function Recalls settings from a saved setup.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage setup.recall(id)
id The setup ID.
Remarks • If a number (n), this parameter is interpreted as a setup number, and the setup is restored from
internal memory. If it is a string, this parameter is interpreted as a path and file name, and the
setup is restored from the corresponding file on the memory stick. The path may be absolute or
relative to the current working directory.
•For an n setting of 0, the unit recalls the factory default (reset) setup. For an n setting of 1 to 5,
the unit recalls a user saved setup from the internal nonvolatile memory.
• Saved setup settings can also be loaded from a USB memory stick. If id is a string, the setup
will be recalled from the memory stick and id is the file name as described above.
Details See User setup in Section 3.
Example Recalls the user-setup at location 2:
setup.recall(2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals