EasyManuals Logo
Home>Keithley>Measuring Instruments>2600a series

Keithley 2600a series User Manual

Keithley 2600a series
696 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
2600AS-901-01 Rev. B / September 2008 Return to Section Topics 3-17
Series 2600A System SourceMeter® Instruments Reference Manual Section 3: Basic Operation
Power measurements
Power calculations
Power readings are calculated from the sourced and measured current or voltage as follows:
P = V × I
Where: P is the calculated power
V is the sourced or measured voltage
I is the measured or sourced current
Basic power measurement procedure
Perform the following steps to perform power measurements. The following procedure assumes
that the SourceMeter instrument is already connected to the DUT as explained in
Section 19.
WARNING Hazardous voltages may be present on the output and guard
terminals. To prevent electrical shock that could cause injury or
death, NEVER make or break connections to the Series 2600A
while the output is on. Power off the equipment from the front
panel or disconnect the main power cord from the rear of the
SourceMeter instrument before handling cables connected to the
outputs. Putting the equipment into standby does not guarantee
the outputs are not powered if a hardware or software fault occurs.
1. For the Model 2602A/2612A/2636A, press the DISPLAY key to select the single-channel
display mode.
2. Set source function and value. Press SRC to select the voltage or current source function
as required, then set the output voltage or current to the desired value. See Step 1 of
Front
panel source-measure procedure earlier in this section.
3. Press the LIMIT key, and set the voltage or current limit high enough for the expected
voltage or current across the DUT to be measured. See Step 2 of
Front panel source-
measure procedure earlier in this section.
4. Press the MEAS or MODE key to display power.
5. Turn on the output, then note the reading on the display. If necessary, press the TRIG key to
display continuous readings.
6. Turn off the output when finished.
Remote power programming
The following paragraphs summarize basic commands necessary for remote power programming
and also give a programming example for a typical power measurement situation.
Remote power command
Use the following command to obtain a power reading:
reading = smuX.measure.p()
See Table 3-6 for more commands necessary to set up source and measure functions and also
Section 19 for more details.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals