2600AS-901-01 Rev. B / September 2008 Return to Section Topics C-21
Series 2600A System SourceMeter® Instruments Reference Manual Appendix C: Status Model
Measurement Event Registers
As shown in Figure C-8, there are eight register sets associated with Measurement Event Status.
Commands are summarized in the Status register sets. Keep in mind that bits can also be set by
using numeric parameter values. For details, see Programming enable and transition registers.
For example, either of the following commands will set the VOLTAGE_LIMIT enable bit:
status.measurement.enable = status.measurement.VOLTAGE_LIMIT
status.measurement.enable = 1
When reading a register, a numeric value is returned. The binary equivalent of this value indicates
which bits in the register are set. For details, see
Reading registers. For example, the following
command will read the measurement enable register:
print(status.measurement.enable)
This register set feeds to bit B0 (MSB) of the Status Byte. The bits used in the Measurement Event
Registers are described as follows:
• Bit B0, Voltage Limit (VLMT): Set bit indicates that the voltage limit was exceeded. This bit
will be updated only when (1) a measurement is taken or (2) the
smuX.source.compliance command is invoked.
• Bit B1, Current Limit (ILMT): Set bit indicates that the current limit was exceeded. This bit
will be updated only when (1) a measurement is taken or (2) the
smuX.source.compliance command is invoked.
• Bit B7, Reading Overflow (ROF): Set bit indicates that an overflow reading has been
detected.
• Bit B8, Buffer Available (BAV): Set bit indicates that there is at least one reading stored in
either or both of the nonvolatile reading buffers.
• Bit B11, Output Enable (OE): Set bit indicates that output enable has been asserted.
• Bit B13, Instrument Summary (INST): Set bit indicates that a bit in the measurement
instrument summary register is set.
For more information on the Measurement Event Registers, refer to Status register sets and the
charts in this appendix.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals