4-8 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 4: Source-Measure Concepts Series 2600A System SourceMeter® Instruments Reference Manual
Figure 4-3A shows the output characteristics for the I-Source. As shown, the Series 2601A and
2602A can output up to 1.01A at 40V, or 3.03A at 6V. Note that when sourcing more than 1.01A,
voltage is limited to 6V.
Figure 4-3B shows the limit lines for the I-Source. The current source limit line represents the
maximum source value possible for the presently selected current source range. The voltage
compliance limit line represents the actual compliance that is in effect (see
Compliance limit).
These limit lines are boundaries that represent the operating limits of the SourceMeter instrument
for this quadrant of operation. The operating point can be anywhere inside (or on) these limit lines.
The limit line boundaries for the other quadrants are similar.
Figure 4-3
Model 2601A/2602A I-Source boundaries
Voltage
Compliance
Limit Line
Current Source
Limit Line
V Measure
40V
6V
1.01A
3.03A
A) Output Characteristics
B) Limit Lines
I Source
Source I
Limit V

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals