19-96 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
reset
printnumber
Function Prints numbers using the format selected for printing reading buffers.
TSP-Link
accessibility
This function cannot be accessed from a remote TSP-Link node.
Usage
There are multiple ways to use this function, depending on how many numbers are to be printed:
printnumber(v1)
printnumber(v1 ,v2)
printnumber(v1 ,v2, ..., vn)
v1, v2, ..., vn Numbers to print.
Remarks This function will print the given numbers using the data format specified by format.data and
other associated attributes.
At least one number must be given.
Also see printbuffer, format.data
Example
Prints three measurements that were previously performed:
format.data = format.ASCII
printnumber(i, v, t)
Example of returned data (i, v, t):
1.02345E-04, 8.76542E-02, 5.29372E-01
reset
Function Resets the logical instruments to the default settings.
TSP-Link
accessibility
This function cannot be accessed from a remote TSP-Link node.
Usage reset()
reset(system)
system Flag indicating what part of the system to reset. Can
be set to true or false.
Remarks •If system is set to true and the node is the master, the entire system will be reset.
•If system is set to false, only the local group will be reset.
The default value of system when no value is given is true.
This function resets all logical instruments in the system or local group. This function is
equivalent to iterating over all the logical instruments and calling the reset method of each.
Resetting the entire system is only permitted when the node is the master. If the node is not the
master, it generates an error.
Also see localnode.reset, smuX.reset

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
CategorySource Measure Unit (SMU)
Model2600A Series
TypePrecision SourceMeter
Voltage Range±200V
Current Range±10fA to ±10A (depending on model)
DC Current±10A
Pulse Current±100A
Maximum Power20W to 200W (depending on model)
Power per Channel20W
Accuracy0.012% basic
InterfaceGPIB, USB, Ethernet
Power Supply100V to 240V, 50Hz to 60Hz
Weight (1-channel)4.5kg
Channels1 or 2
Dimensions (1-channel)89 mm high x 214 mm wide x 429 mm deep (3.5 in x 8.4 in x 16.9 in)
Dimensions (2-channel)89 mm high x 214 mm wide x 429 mm deep (3.5 in x 8.4 in x 16.9 in)

Related product manuals