2600AS-901-01 Rev. B / September 2008 Return to Section Topics 19-9
Series 2600A System SourceMeter® Instruments Reference Manual Section 19: Remote Commands
2 Two
3 Three
Counting from one to four,
in steps of two:
1 One
3 Three
Counting elements in list
on numeric index
1 One
2 Two
3 Three
4 Four
5 Five
6 Six
Counting elements in list
using repeat
1 One
2 Two
3 Three
4 Four
5 Five
6 Six
Command programming notes
Conventions
For the following command reference, it is necessary to understand the following conventions:
Wild characters
Many source-measure unit (SMU) commands are expressed in a generic form using wild
characters. A wild character indicates a SMU channel, function, or array index. Keep in mind that
wild characters used in the generic form are NOT to be included in the command sent to the
instrument.
X and Y
The X character is used for functions and attributes to indicate the SMU channel (a or b) and Y is
used to indicate the SMU function (
v, i, r, or p). For example, the attribute for the source output
setting is generically expressed as follows:
smuX.source.levelY
To program SMU Channel A to 5 Volts, the following command statement is to be sent to the
instrument:
smua.source.levelv = 5.0
To program SMU channel B to 1 mA, the following command statement is to be sent to the
instrument:
smub.source.leveli = 0.001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals