2600AS-901-01 Rev. B / September 2008 Return to Section Topics C-19
Series 2600A System SourceMeter® Instruments Reference Manual Appendix C: Status Model
Figure C-11
Standard event register
Operation Event Registers
As shown in Figure C-3, there are 22 register sets associated with Operation Event Status.
Commands are summarized in the Status register sets. Keep in mind that bits can also be set by
using numeric parameter values. For details, see Programming enable and transition registers.
For example, either of the following commands will set the CAL enable bit (B0):
status.operation.enable = status.operation.CAL
status.operation.enable = 1
When reading a register, a numeric value is returned. The binary equivalent of this value indicates
which bits in the register are set. For details, see
Reading registers. For example, the following
command will read the operation enable register:
print(status.operation.enable)
Table C-4
Standard event commands
Command Description
*ESR?
or
print(status.standard.event)
Read Standard Event Status Register.
*ESE <mask>
or
status.standard.enable = <mask>
Program the Event Status Enable Register:
<mask> = 0 to 255
See Status register sets.
*ESE?
or
print(status.standard.enable)
Read Event Status Enable Register.
DDE
(B3)
QYE
(B2)
î™–
(B1)
OPC
(B0)
EXE
(B4)
CME
(B5)
URQ
(B6)
PON
(B7)
î™–
(B15 - B8)
Standard
Event Status
Enable
Register
&
&
&
&
&
&
OR
To Event
Summary Bit
(ESB) of Status
Byte Register.
(See Figure D-1)
PON = Power On
URQ = User Request
CME = Command Error
EXE = Execution Error
DDE = Device-Dependent Error
*ESE
*ESE?
&
DDE
(B3)
QYE
(B2)
î™–
(B1)
OPC
(B0)
EXE
(B4)
CME
(B5)
URQ
(B6)
PON
(B7)
î™–
(B15 - B8)
Standard
Event Status
Register
*ESR?
QYE = Query Error
OPC = Operation Complete
& = Logical AND
OR = Logical OR
print(status.standard.event)
status.standard.enable
print(status.standard.enable)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals