19-102 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
setup.save
Function Saves the present setup as a user-setup.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage setup.save(id)
id The setup ID.
Remarks • Numbers 1 through 5 are used to designate user-setup locations. When you save to one of these
locations, the previous setup at that location is overwritten.
• If this parameter is a number, it is interpreted as a setup number, and the setup is saved to
internal memory. If it is a string, this parameter is interpreted as a path and file name, and the
setup is saved to the corresponding file on the memory stick. The path may be absolute or
relative to the current working directory.
Details See User setup in Section 3.
Example Saves the present setup at location 5:
setup.save(5)