20-10 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 20: Calibration Series 2600A System SourceMeter® Instruments Reference Manual
Figure 20-1
Connections for voltage calibration
c. Perform each calibration step listed in Table 20-2 (Model 2601A/2602A),or Table 20-3
(Model 2611A/2612A), or (Model 2635A/2636A) as follows:
1) Select the range being calibrated with this command:
smua.source.rangev = range
CHANNEL B
SENSE
LO
LO
GUARD
SENSE
HI
HI
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSP-Link
LINE FUSE
SLOWBLOW
3.15A, 250V
!
A LOB LO
R
CHANNEL A
SENSE
LO
LO
GUARD
SENSE
HIHI
CHANNEL B
SENSE
LO
LO
GUARD
SENSE
HI
HI
Model 2636 channel A
connections shown
LAN
Input HI
Input LO
S LO
LO
HI
S HI
Connections Shown
2636A Channel A
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-23
2
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
SSLO
LOHI
HIGGG G
CHANNEL
B
!
Digital Multimeter
Input HI
Input LO
S HI
HI
S LO
LO
2602A/2612A Channel A
Connections Shown