Section 20: Calibration Series 2600A System SourceMeter® Instruments Reference Manual
Step 7. Lock out calibration
To lock out further calibration, send the following command after completing the calibration
procedure:
smua.cal.lock()
Step 8. Repeat calibration procedure for Model 2602A/2612A/2636A Channel B
For the Models 2602A, 2612A, and 2636A only, repeat the entire procedure above for Channel B.
Be sure to:
• Make test connections to Channel B terminals.
• Substitute “smub” for “smua” in all commands.