3-8 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 3: Basic Operation Series 2600A System SourceMeter® Instruments Reference Manual
2. Move the cursor to the digit to change, then press the navigation wheel to enter the EDIT
mode, as indicated by the EDIT indicator.
3. Use the RANGE keys to select a range that will accommodate the value you want to set.
(See
Section 6 for range information.) For best accuracy, use the lowest possible source
range.
4. Enter the desired source value, then press ENTER or the navigation wheel to complete
editing.
Step 2: Set compliance limit.
Perform the following steps to edit the compliance limit value:
1. For the Model 2601A/2611A/2635A or the Model 2602A/2612A/2636A single-channel
display mode, press the LIMIT key.
2. For the Model 2602A/2612A/2636A dual-channel display mode, press CONFIG then LIMIT,
then select CURRENT or VOLTAGE. Press ENTER or the navigation wheel.
3. Move the cursor to the digit to change, then press the navigation wheel to enter the EDIT
mode, as indicated by the EDIT indicator.
4. Enter the desired limit value, then press ENTER or the navigation wheel to complete
editing.
Step 3: Select measurement function and range.
Select measurement function and range as follows:
1. Put the Model 2602A/2612A/2636A in the single-channel display mode, then select the
desired measurement function by pressing MEAS or MODE.
2. Select the desired measurement range with the RANGE keys, or enable AUTO RANGE,
keeping the following points in mind:
• When measuring the source (such as Source V Measure V), you cannot select the
measurement range using the RANGE keys. The selected source range determines the
measurement range.
• When not measuring the source (such as Source V Measure I), measurement range
selection can be done manually or automatically. When using manual ranging, use the
lowest possible range for best accuracy. In auto range, the Series 2600A automatically
goes to the most sensitive range to make the measurement.
Step 4: Turn output on.
Turn the output on by pressing the ON/OFF OUTPUT key. The OUTPUT indicator light will turn on.
Step 5: Observe readings on the display.
Observe the readings on the display. Press TRIG if necessary to trigger the unit to begin taking
readings. For the single-channel display mode, the readings will appear on the top line, while
source and limit values are on the bottom line.
Step 6: Turn output off.
When finished, turn the output off by pressing the ON/OFF OUTPUT key. The OUTPUT indicator
light will turn off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals