2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-19
Figure 6-7
Model 2425/2430 operating boundaries (T
amb
30°C)
–I
= 100% Duty Cycle
= 60% Duty Cycle
(III)
Source
–V
+V
+I
(IV)
Sink
(II)
Sink
(I)
Source
-20V 20V
-3A
-2A
-1A
1A
3A
2A
0.5A
-0.5A
-100V 100V
10A
6A
-6A
-10A
-100V 100V
= 8% Duty Cycle
= 5% Duty Cycle
+I
-I
+V-V
A. Model 2425 and Model 2430 DC Mode
B. Model 2430 Pulse Mode
(IV)
Sink
(I)
Source
(II)
Sink
(III)
Source