EasyManuals Logo
Home>Keithley>Measuring Instruments>SourceMeter 2400

Keithley SourceMeter 2400 User Manual

Keithley SourceMeter 2400
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #361 background imageLoading...
Page #361 background image
15-14 Status Structure 2400 Series SourceMeter
®
User’s Manual
Bit B7, Reading Overflow (ROF) — Set bit indicates that the volts or amps read-
ing exceeds the selected measurement range of the SourceMeter.
Bit B8, Buffer Available (BAV) — Set bit indicates that there are at least two read-
ings in the buffer.
Bit B9, Buffer Full (BFL) — Set bit indicates that the trace buffer is full.
Bit B10 — Limit 4 - Contact Check (see Appendix F).
Bit B11, Interlock Asserted (Int) — Set bit indicates that the interlock line is at
digital low (asserted). The source output can be turned on.
Bit B12, Over Temperature (OT) — Set bit indicates that an over temperature
condition exists. The source output cannot be turned on.
Bit B13, Over Voltage Protection (OVP) — Set bit indicates that the source is
being limited at the programmed limit level.
Bit B14, Compliance (Comp) — Set bit indicates that the source is in compliance.
Bit B15 — Not used.
Figure 15-6
Measurement event status
LL3
(B3)
HL2
(B2)
LL2
(B1)
L1
(B0)
HL3
(B4)
LP
(B5)
RAV
(B6)
ROF
(B7)
BAV
(B8)
BFL
(B9)
CC
(B10)
OT
(B12)
Measurement
Condition Register
BFL
(B9)
CC
(B10)
Measurement Event
Register
CC
(B10)
Measurement Event
Enable Register
&
&
&
&
&
&
OR
To Measurement
Summary Bit
(MSB) of Status
Byte Register.
(See Figure 15-3.)
Comp = In Compliance
OVP = Over Voltage protection acgentcy
OT = Over temperature
INT = Interlock Asserted
CC = Contact Check
BFL = Buffer Full
BAV = Buffer Available
ROF = Reading Overflow
RAV = Reading Available
INT
(B11)
&
&
&
&
LP = Limits Pass
HL3 = High Limit 3
LL3 = Low Limit 3
HL2 = High Limit 2
LL2 = Low Limit 2
L1 = Limit 1
& = Logical AND
OR = Logical OR
OVP
(B13)
Comp
(B14)
(B15)
INT
(B11)
OT
(B12)
OVP
(B13)
Comp
(B14)
(B15)
INT
(B11)
OT
(B12)
OVP
(B13)
Comp
(B14)
(B15)
&
&
&
&
stat:meas:cond?
stat:meas?
stat:meas:enab <NRF>
stat:meas:enab?
&
L1
(B0)
LL2
(B1)
HL2
(B2)
LL3
(B3)
HL3
(B4)
LP
(B5)
RAV
(B6)
ROF
(B7)
BAV
(B8)
BFL
(B9)
BAV
(B8)
ROF
(B7)
RAV
(B6)
LP
(B5)
HL3
(B4)
LL3
(B3)
HL2
(B2)
LL2
(B1)
L1
(B0)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 and is the answer not in the manual?

Keithley SourceMeter 2400 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400
CategoryMeasuring Instruments
LanguageEnglish