2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-29
Figure 6-14
V-Source operating examples
Current Limit
Load Line
Operating
Point
Voltage Source
Load Line
V-Source (V
S
)
I-Meter
(I
M
)
25mA
50mA
50V
2k DUT Load Line (R)
I
M
= V
S
/ R
= 50V/2k
= 25mA
A. Normal V-source operation
Current Limit
Load Line
Operating
Point
Voltage Source
Load Line
V-Source (V
S
)
I-Meter
(I
M
)
50mA
50V
V
S
= I
M
· R
= (50mA) (800)
= 40V
B. V-Source in compliance
40V
800 DUT Load Line (R)