18-96 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
RECall <NRf>
:SOURCe:MEMory:RECall <NRf> Return to specified setup
Parameters <NRf> = 1 to 100 Specify memory location
Description This command is used to return the SourceMeter to the setup stored at
the specified memory location.
Set scaling factor
TRIGgered:SFACtor <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor <n> Set current scaling factor
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor <n> Set voltage scaling factor
Parameters <n> = -999.9999e+18 to 999.9999e+18 Scaling factor
Query :SFACtor? Query scaling factor
Description :SFAC instructs the SourceMeter to source the scaling factor times the
previous source memory location value. For example, if 10.0V is stored
in the first source memory (Source I, Measure V Mode), and the unit is
in the Source V, Measure I mode with :SFAC set to 0.1 and enabled, the
unit will output 1.0V for the second source memory location.
NOTE These commands work only with source memory sweeps.
TRIGgered:SFACtor:STATe <b>
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor:STATe <b> Enable/disable current scaling
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor:STATe <b> Enable/disable voltage scaling
Parameters <b> = 1 or ON Enable scaling
0 or OFF Disable scaling
Query :SFACtor:STATe? Query enabled/disabled scaling state
Description :SFAC:STAT enables or disables scaling.
NOTE These commands work only with source memory sweeps.
Sweep and list program examples
Linear voltage sweep
Linear voltage sweep from 1V to 10V in 1V increments:
*RST
SOUR:FUNC:MODE VOLT
SOUR:SWE:SPAC LIN