User’s Manual SCPI Command Reference 18-121
WARNING With auto output-off disabled, the source output will remain on after
all programmed source-measure operations are completed. Beware of
hazardous voltage (≥30VDC, 42.4 peak-to-peak) that may be present
on the output terminals.
With auto output-off enabled, an initiation will start operation immedi-
ately. The source output will automatically turn on at the beginning of
each SDM (source-delay-measure) cycle and turn off after each mea-
surement is completed.
Abort source/measure cycle
ABORt Abort operation
Description When this action command is sent, the SourceMeter aborts operation
and returns to the idle state.
A faster way to return to idle is to use the DCL or SDC command.
With auto output-off enabled (:SOURce1:CLEar:AUTO ON), the output
will remain on if operation is terminated before the output has a chance
to automatically turn off.
Program trigger model
COUNt <n>
:ARM[:SEQuence[1]][LAYer[1]]:COUNt <n> Set arm count
:TRIGger[:SEQuence[1]]:COUNt <n> Set trigger count
Parameters <n> = 1 to 2500 Specify count (see NOTE)
DEFault Sets count to 1
MINimum Sets count to 1
MAXimum See NOTE
INFinite (ARM:COUNt only)
NOTE The product of arm count and trigger count cannot exceed 2500.
Query :COUNt? Queries programmed count
:COUNt? DEFault Queries *RST default count
:COUNt? MINimum Queries lowest allowable count
:COUNt? MAXimum Queries largest allowable count
Description This command is used to specify how many times an operation is per-
formed in the specified layer of the trigger model.