1-26 Getting Started 2400 Series SourceMeter
®
User’s Manual
Front panel control
Front panel display circuitry is controlled from the DISABLE DISPLAY configuration
menu, which is accessed by pressing CONFIG and then EDIT (or TOGGLE). To select an
option (NOW, NEVER, SWEEP, or STORE), use the EDIT cursor keys to place the cursor
on the desired option, then press ENTER.
The options for DISABLE DISPLAY are explained as follows:
NOW — Select this option to disable the display now.
NEVER — Select this option if you do not want the display to disable.
SWEEP — Select this option if you want the display to disable while performing a sweep.
The display will disable as soon as sweep is started. The display will automatically re-
enable after the sweep is completed.
STORE — Select this option if you want the display to disable when storing source-
measure readings in the buffer. The display will disable as soon as the buffer is enabled.
The display will automatically re-enable after the storage process is completed. Note that
with this option, the display will disable while performing a sweep. Sweep readings are
automatically stored in the buffer.
Remote command programming
Use the following SCPI commands to enable or disable the front panel display circuitry:
:DISPlay:ENABle OFF Disable the display
:DISPlay:ENABle ON Enable the display
Configuration menus
There are a number of configuration menus that can be accessed by pressing the CONFIG
key followed by the appropriate function or mode key. For example, you can configure the
voltage source by pressing CONFIG then SOURCE V. Configuration menus, which are
summarized in Table 1-6 through Table 1-10, are available for the following operating
modes:
Measure functions (V, , FCTN): Table 1-6.
SOURCE V, SOURCE I, and RANGE: Table 1-7.
REL, FILTER, and LIMIT: Table 1-8.
TRIG: Table 1-9.
SWEEP, DIGITS, SPEED, and STORE: Table 1-10.
ON/OFF OUTPUT and Display (EDIT or TOGGLE): Table 1-11.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 and is the answer not in the manual?

Keithley SourceMeter 2400 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals