18-126 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
OUTPut <event list>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OUTPut <event list> Arm layer events
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut <event list> Trigger layer events
Parameters Arm Layer Triggers
<event list >: TENTer Trigger on entering trigger layer
TEXit Trigger on exiting trigger layer
NONE Disable arm layer output trigger
Trigger Layer Triggers
<event list>: SOURce Output trigger after source level is set
DELay Output trigger after delay period
SENSe Output Trigger after measurement
NONE Disable trigger layer triggers
NOTE Each event in the list must be separated by a comma. (i.e., :arm:output source,
delay, sense).
Query :OUTPut? Query output trigger event(s)
Description This command is used to specify when trigger pulses occur on the spec-
ified output trigger line of the Trigger Link (see “OLINe <NRf>,”
page 18-125).
Arm Layer Triggers — With TEXit selected, an output trigger will
occur when exiting the trigger layer. With TENTer selected, an output
trigger will occur when entering the trigger layer. With NONE selected,
the arm layer output trigger is disabled.
Trigger Layer Triggers — You can specify from one to all three events.
Each event in the list must be separated by a comma (,).
The SOURce, DELay, and MEASure events refer to the Source-Delay-
Measure (SDM) cycle. This is the Device Action in the Trigger Model.
(See Figure 11-8 and Figure 11-9.) With SOURce specified, an output
trigger occurs after the source is set. With DELay specified, an output
trigger occurs after the delay period. With MEASure specified, an out-
put trigger occurs after the measurement.
NOTE When disabling triggers, the NONE parameter must be sent alone (i.e., trig:outp
none). If it is listed with any of the other event parameters, NONE will be
ignored.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com