18-4 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
:CALCulate2
:LIMit[1] Path to control LIMIT 1 test: ✓
:COMPliance Configure Limit 1 test:
:FAIL <name> Specify fail condition (IN or OUT of
compliance).
IN ✓
:FAIL? Query fail condition.
:SOURce2 <NRf>
| <NDN>
Specify output fail pattern (0 to 7,
3-bit; 0 to 15, 4-bit).
15 ✓
:SOURce2? Query fail bit pattern.*
:STATe <b> Enable or disable Limit 1 test. OFF ✓✓
:STATe? Query state of Limit 1 test. ✓
:FAIL? Returns result of Limit 1 test: 0 (pass) or
1 (fail).
✓
:LIMit2 Path to control LIMIT 2 test: ✓
:UPPer Configure upper limit: ✓
[:DATA] <n> Specify upper limit (-9.999999e20 to
9.999999e20).
1 ✓✓
[:DATA]? Query upper limit. ✓
:SOURce2 <NRf>
| <NDN>
Specify output fail pattern for grading
mode (0 to 7, 3-bit; 0 to 15, 4-bit).
15 ✓
:SOURce2? Query fail bit pattern.*
:LOWer Configure lower limit: ✓
[:DATA] <n> Specify lower limit (-9.999999e20 to
9.999999e20).
-1 ✓✓
[:DATA]? Query lower limit. ✓
:SOURce2 <NRf>
| <NDN>
Specify output fail pattern for grading
mode (0 to 7, 3-bit; 0 to 15, 4-bit).
15 ✓
:SOURce2? Query fail bit pattern.*
:PASS Path to specify pass pattern for sorting
mode:
:SOURce2 <NRf>
| <NDN>
Specify output pass pattern (0 to 7,
3-bit; 0 to 15, 4-bit).
15 ✓✓
:SOURce2? Query pass bit pattern.*
:STATe <b> Enable or disable Limit 2 test. OFF ✓✓
:STATe? Query state of Limit 2 test. ✓
:FAIL? Return result of Limit 2 test: 0 (pass) or
1 (fail).
✓
Table 18-1 (continued)
CALCulate command summary
Command Description
Default
parameter SCPI
Source
memory