Channel Setup
- 18 -
MXD70 Electrodeless Conductivity
Setup and Operating Guide
Channel Setup
Temperature Compensation
Temperature compensation is enabled by setting
this to “In”.
/ – Select Option
EXIT
– Cancel
– Save Selection
Temperature Compensation Base
Sets the temperature compensation base. See
Appendix B - Temperature Coefficient for more
information. Only Available if Temperature
Compensation is set to in.
/ – Select Option
EXIT
– Cancel
– Save Selection
Temperature Compensation Slope
Sets the temperature compensation slope. See
Appendix B - Temperature Coefficient for more
information. Only Available if Temperature
Compensation is set to in.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value
Temperature Compensation Mode
To use temperature compensation from a measured
input select “Auto”, else select “Manual” to enable a
fixed value entry. Only Available if Temperature
Compensation is set to in.
/ – Select Option
EXIT
– Cancel
– Save Selection