Appendix
MXD70 Dissolved Oxygen
Setup and Operating Guide
- 41 -
Appendix
Appendix B - Temperature Data
The table below lists approximate resistance values of temperature sensors that may be used with the
MXD70 series. Not all options are available on all input types.
Temperature
( C )
PT1000
RTD
PT100
RTD
1K Thermistor
BJ 22K
Thermistor
0
1000.0 100.00 2691 64.88 k
10
1039.0 103.90 1779 41.34 k
20
1077.9 107.79 1204 26.97 k
25
1097.3 109.73 1000 22.00 k
30
1116.7 111.67 833.7 18.03 k
40
1155.4 115.54 589.0 12.30 k
50
1194.0 119.40 423.9 8.57 k
60
1232.4 123.24 310.5 6.07 k
70
1270.7 127.07 231.0 4.38 k
80
1308.9 130.89 174.5 3.21 k
90
1347.0 134.70 133.6 2.39 k
100
1385.0 138.50 103.6 1.80 k