Profibus
92 Rev. G Nov/2015 MI3
14.2 Programming
14.2.1 Parameter Data
Byte
Address
without
offset
Description
Format
Range
0 to 6
Fix
7
DP-V1 Status1
8
DP-V1 Status2
9
DP-V1 Status3
10
3
Temp. unit
67=°C, 70=°F
67 or 70
11
4
Reserved
12, 13
5
Bottom temp. of output 1
in 0.1°C /°F
14, 15
7
Top temp. of output 1
in 0.1°C /°F
16, 17
9
Bottom temp. of output 2
in 0.1°C /°F
18, 19
11
Top temp. of output 2
in 0.1°C /°F
20
13
Source (head) for output-1
head number
1…8
21
14
Type of source for output-1
73 = I (internal temp.)
84 = T (object temp.)
73 or 84
22
15
Analog output mode 1
5 = TCJ
6 = TCK
7 = TCR
8 = TCS
9 = 0…5V
10 = 0…10V
99 = tristate (disabled)
5…10, 99
default: 9
23
16
Source (head) for output-2
head number
1…8
24
17
Type of source for output-2
73 = I (internal temp.),
84 = T (object temp.)
73 or 84
25
18
Analog output mode 2
0 = 0…20mA
4 = 4…20mA
9 = 0…5V
10 = 0…10V
99 = tristate (disabled)
0, 4, 9, 10, 99
default: 4
26…39
reserved
40…43
reserved
44, 45
37
Emissivity Head_1
* 1000 (0.9 900)
100 … 1100
46, 47
39
Transmissivity Head_1
* 1000 (1.0 1000)
100 … 1000
48, 49
41
Averaging time Head_1
* 0.1s (1s 10)
0 …9990
50, 51
43
Peak hold time Head_1
* 0.1s (1s 10)
0 …9990
52, 53
45
Valley hold time Head_1
* 0.1s (1s 10)
0 …9990
54, 55
47
Ambient temp. Head_1
in °C /°F
dev. range min..
max
56, 57
49
Setpoint relay Head_1
in °C /°F
dev. range min..
max
58
51
Relay alarm output control Head_1
0 = off
1 = target temp.
2 = internal temp.
0, 1 or 2