2 Command System RIGOL
DS2000E Programming Guide
:MASK:SOURce
Syntax
:MASK:SOURce <source>
:MASK:SOURce?
Description
Set or query the measurement source of the pass/fail test.
Parameter
Name Type Range Default
<source> Discrete {CHANnel1|CHANnel2} CHANnel1
Explanation
Only the channel enabled can be selected as the measurement source of the pass/fail test.
Return Format
The query returns CHAN1 or CHAN2.
:MASK:OPERate
Syntax
:MASK:OPERate <oper>
:MASK:OPERate?
Description
Run or stop the pass/fail test; or query the current status of the pass/fail test.
Parameter
Name Type Range Default
<oper> Discrete {RUN|STOP} STOP
Explanation
Before executing this command, use the
:MASK:ENABle command to enable the pass/fail test.
Return Format
The query returns RUN or STOP.