2 Command System RIGOL
DS2000E Programming Guide
2-147
:MEASure:HISTory:DISPlay
Syntax
:MEASure:HISTory:DISPlay <bool>
:MEASure:HISTory:DISPlay?
Description
Enable or disable the measurement history; or query the current on/off status of the measurement
history.
Parameter
Name Type Range Default
<bool> Bool {{1|ON}|{0|OFF}} 0|OFF
Explanation
The measurement history cannot be enabled when there is not any measured item.
When the measurement history is enabled, the latest 10 measurement results of at most 5
measurement items turned on last can be displayed.
Return Format
The query returns 1 or 0.
:MEASure:HISTory:DMODe
Syntax
:MEASure:HISTory:DMODe <mod>
:MEASure:HISTory:DMODe?
Description
Set or query the display mode of the history measurement data.
Parameter
Name Type Range Default
<mod> Discrete {TABLe|GRAPh} GRAPh
Explanation
TABLe: display the measurement results of the last 10 measurements of at most 5 measurement
items that are enabled last in table mode.
GRAPh: display the measurement results of the last 10 measurements of at most 5 measurement
items that are enabled last in graph mode. The measurement points are connected using linear
interpolation.
Return Format
The query returns TABL or GRAP.

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals