EasyManuals Logo
Home>Rigol>Test Equipment>DS2000E Series

Rigol DS2000E Series Programming Guide

Rigol DS2000E Series
365 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #256 background imageLoading...
Page #256 background image
RIGOL 2 Command System
2-240 DS2000E Programming Guide
:TIMebase[:MAIN]:OFFSet
Syntax
:TIMebase[:MAIN]:OFFSet <offset>
:TIMebase[:MAIN]:OFFSet?
Description
Set or query the horizontal offset of the main time base and the unit is s.
Parameter
Name Type Range Default
<offset> Real
The range of <offset> is related to the horizontal time base mode (refer
to the
:TIMebase:MODE) and the operating state of the oscilloscope.
Y-T mode
RUN: When TimeScale is less than 10ms/div, the range of <offset>
is from (-MemDepth/SampleRate) to 1s.
When TimeScale is greater than or equal to 10ms/div, the
range of <offset> is from (-MemDepth/SampleRate) to (100 x
TimeScale).
STOP: -7000s to 7000s
Roll mode
RUN: This command is not available.
STOP: -7000s to 0
0
Note:
For the TimeScale, refer to the
:TIMebase[:MAIN]:SCALe command.
For the MemDepth, refer to the
:ACQuire:MDEPth command.
For the SampleRate, refer to the
:ACQuire:SRATe? command.
Return Format
The query returns the horizontal offset of the main time base in the scientific notation.

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals