RIGOL 2 Command System
2-318 DS2000E Programming Guide
:WAVeform:MODE
Syntax
:WAVeform:MODE <mode>
:WAVeform:MODE?
Description
Set or query the waveform data reading mode.
Parameter
Name Type Range Default
<mode> Discrete {NORMal|RAW|MAXimum} NORMal
Explanation
NORMal: read the waveform data currently displayed on the screen.
RAW: read the waveform data from the internal memory.
Note: The waveform data in the internal memory can only be read when the oscilloscope is in the
stop state and please do not operate the oscilloscope during the reading process.
MAXimum: in the run state, read the waveform data displayed on the screen; in the stop state, read
the waveform data from the internal memory.
When the channel source of waveform data reading (refer to the
:WAVeform:SOURce command) is
set to MATH or FFT, only the waveform data displayed on the screen can be read.
Return Format
The query returns NORM, RAW, or MAX.