2 Command System RIGOL
DS2000E Programming Guide
:MEASure:RPHase:SMAXimum
:MEASure:RPHase:SMINimum
:MEASure:RPHase:SCURrent
:MEASure:RPHase:SAVerage
:MEASure:RPHase:SDEViation
Syntax 1
:MEASure:RPHase:SMAXimum <chanA>,<chanB>
:MEASure:RPHase:SMINimum <chanA>,<chanB>
:MEASure:RPHase:SCURrent <chanA>,<chanB>
:MEASure:RPHase:SAVerage <chanA>,<chanB>
:MEASure:RPHase:SDEViation <chanA>,<chanB>
Description 1
Enable the phase deviation (rising edge-rising edge) measurement function between the two specified
channels and the statistic function of the measurement results.
Syntax 2
:MEASure:RPHase:SMAXimum? <chanA>,<chanB>
:MEASure:RPHase:SMINimum? <chanA>,<chanB>
:MEASure:RPHase:SCURrent? <chanA>,<chanB>
:MEASure:RPHase:SAVerage? <chanA>,<chanB>
:MEASure:RPHase:SDEViation? <chanA>,<chanB>
Description 2
Query the statistic maximum value (SMAXimum), minimum value (SMINimum), current value (SCURrent),
average value (SAVerage), or standard deviation (SDEViation) of the phase deviation (rising edge-rising
edge) measurement results when the phase deviation (rising edge-rising edge) measurement function
between the two specified channels and the statistic function of the measurement results are enabled.
Parameter
Name Type Range Default
<chanA> Discrete {CHANnel1|CHANnel2} --
<chanB> Discrete {CHANnel1|CHANnel2} --
Return Format
The query returns the statistic maximum value, minimum value, current value, average value, or standard
deviation of the phase deviation (rising edge-rising edge) measurement results in scientific notation.
When the measurement results are invalid, the query returns 9.9e37.