RIGOL 2 Command System
2-260 DS2000E Programming Guide
:TRIGger:DURAtion:WHEN
Syntax
:TRIGger:DURATion:WHEN <when>
:TRIGger:DURATion:WHEN?
Description
Set or query the trigger condition of duration trigger.
Parameter
Name Type Range Default
<when> Discrete {GREater|LESS|GLESs} GREater
Explanation
GREater: you need to specify a time. The oscilloscope triggers when the duration of the pattern is
greater than the preset time.
LESS: you need to specify a time. The oscilloscope triggers when the duration of the pattern is lower
than the preset time.
GLESs: you need to specify an upper limit of the time and lower limit of the time. The oscilloscope
triggers when the duration of the pattern is lower than the preset upper limit of the time and greater
than the preset lower limit of the time.
Return Format
The query returns GRE, LESS, or GLES.