2 Command System RIGOL
DS2000E Programming Guide
2-273
:TRIGger:PATTern:PATTern
Syntax
:TRIGger:PATTern:PATTern <pa_ch1>[,<pa_ch2>]
:TRIGger:PATTern:PATTern?
Description
Set or query the pattern of each channel of pattern trigger.
Parameter
Name Type Range Default
<pa_ch1> Discrete {H,L,X,R,F} X
<pa_ch2> Discrete {H,L,X,R,F} X
Explanation
The patterns consist of 2 analog channels.
<pa_ch1> and <pa_ch2> set the patterns of analog channels CH1 and CH2. Users can send 2
parameters to set the pataterns of the two channels. Also, you can omit the parameter <pa_ch2> to
set the pattern of CH1 (the pattern state of CH2 remains unchanged). No matter whether the
parameters are omitted, the query command returns the currently set pattern for all the channels.
H represents high level (higher than the threshold level of the channel). L represents low level (lower
than the threshold level of the channel). X denotes that this channel is ignored (this channel is not
used as a part of the pattern. When all the channels in the pattern are set to X, the oscilloscope will
not trigger).
H: high level (higher than the threshold level of the channel).
L: low level (lower than the threshold level of the channel).
X: ignore this channel (this channel is not used as a part of the pattern). When all the channels in the
pattern are set to X, the oscilloscope will not trigger.
R, F: rising edge, falling edge. In the pattern, you cannot specify multiple channels as rising edge or
falling edge at the same time. If an edge item is already defined and then another edge item is
defined in another channel in the pattern, then the edge item defined formerly will be replaced by X.
Please distinguish "omit parameter" from "ignore channel". The former means that the parameters
corresponding to some channels are not sent when sending the command, and the pattern states of
these channels remain unchanged. The latter means that the channel is set to X when sending the
command, and the state of this channel is ignored when the instrument judges the patterns.
Return Format
The query returns the currently set patterns for all the channels (2).
Example
:TRIGger:PATTern:PATTern H /*Set the pattern of CH1 to H, and the pattern state of CH2 remains
unchanged*/
:TRIGger:PATTern:PATTern? /*The query returns H,X*/

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals