RIGOL 2 Command System
2-304 DS2000E Programming Guide
:TRIGger:TIMeout:SLOPe
Syntax
:TRIGger:TIMeout:SLOPe <slope>
:TRIGger:TIMeout:SLOPe?
Description
Set or query the edge type of timeout trigger.
Parameter
Name Type Range Default
<slope> Discrete {POSitive|NEGative|ALTernate} POSitive
Explanation
POSitive: start timing when the rising edge of the input signal passes through the trigger level.
NEGative: start timing when the falling edge of the input signal passes through the trigger level.
ALTernate: start timing when any edge of the input signal passes through the trigger level.
Return Format
The query returns POS, NEG, or ALT.
:TRIGger:TIMeout:TIMe
Syntax
:TRIGger:TIMeout:TIMe <NR3>
:TRIGger:TIMeout:TIMe?
Description
Set or query the timeout time of timeout trigger.
Parameter
Name Type Range Default
<NR3> Real 16ns to 4s 1us
Return Format
The query returns the timeout time of timeout trigger in scientific notation.

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals