Climatic conditions
Permissible ambient temperature
for OXYMAT 7 and CALOMAT 7
1)
0 ... 50 °C (32 ... 122 °F)
Permissible ambient temperature
during operation for ULTRAMAT 7
5 °C ... +45 °C (41 ... +133 °F)
Pollution degree 2
1) Permissible temperature range is reduced if module is used in combination with Ultramat 7
7.1.2.1 Using Ex p
WARNING
Purging gas failure for operation with type of explosion protection Ex p
Explosion hazard
In the case of operation with the type of protection Ex p, explosive media can form if the
purging gas fails.
De-energize all sources of ignition when purging gas fails.
Stop supply of 󹪞ammable process gases.
WARNING
Using Ex px
Risk of explosion in hazardous areas.
When using type of protection Ex px, route all electric inlets and outlets across Ex coupling
relays.
7.1.2.2 Cable glands
Note
Ranges of the cable glands
Use cables matched to the ranges of the cable glands:
Cable gland M16 x 1.5 (plastic): Range 5 to 10 mm (0.2" .. 0.39")
Cable gland M16 x 1.5 (metal): Range 6 to 10 mm (0.2" .. 0.39")
Cable gland M20 x 1.5 (plastic/metal): Range 10 to 14 mm (0.4" .. 0.55")
Technical data
7.1General technical speci󹪝cations
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB 91

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals