Figure 2: Installing a connector saver
3. Use the backing wrench on the connector saver to prevent damage caused by rotation of the connectors when installing a
connector.
4. Use the supplied torque wrench to tighten (8 inch-pounds) your cable connector to the connector saver.
Figure 3: Installing a cable
CAUTION. Failure to prevent rotation of the center conductors when making connections to the ATI input will damage your
instrument. If necessary use a second backing wrench on the cable, to prevent its rotation. Use the supplied torque wrench
to tighten (8 inch-pounds) all connectors.