Performance Tests
4-40
TDS7104 & TDS7054 Service Manual
b. Initialize the oscilloscope: Press DEFAULT SETUP.
c. Modify the default settings:
H From the tool bar, touch Horiz and select the Acquisition tab.
H Touch Average and set the number of averages to 16.
2. Confirm input channels are within limits for DC accuracy at maximum offset
and position: Do the following substeps — test CH 1 first, skipping substep
2a since CH 1 is already selected from step 1.
a. Select an unchecked channel:
H From the tool bar, touch MEAS and then Clear to remove the
previous measurement.
H Press the Vertical button of the channel just confirmed to remove the
channel from the display.
H Press the front-panel Vertical button that corresponds to the channel
you are to confirm.
H Set the generator output to 0 V.
H Move the test hookup to the channel you selected.
b. Turn on the measurement Mean for the channel:
H From the tool bar, touch MEAS and select the Ampl tab, then touch
Mean to measure the mean of the current channel.
H Touch Setup Statistics and touch Measurement Statistics Mean to
display the measurement statistics of the mean measurement.
H Press Close.
c. Set the vertical scale: Set the vertical S CALE to one of the settings in
Table 4--6 that is not yet checked. (Start with the first setting listed).

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals