EasyManuals Logo
Home>Tektronix>Test Equipment>TDS6000 Series

Tektronix TDS6000 Series Service Manual

Tektronix TDS6000 Series
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
Adjustment Procedures
5-10
TDS7104 & TDS7054 Service Manual
4. Locate the test program file: After the installation is complete, open the
directory C:\tekcats\ and then locate the file tds7000_nt.exe.Thisisthefile
you will double click to execute either the performance verification
procedure or the adjustment procedure. The install procedure also created a
desktop shortcut (link) to this file for your convenience.
Adjustment Program Execution
Once the instrumentation is connected, configured, and has been allowed to
warm up for at least 20 minutes, you can start the adjustment program. To start
the adjustment program, select the file C:\tekcats\tds7000_nt.exe (or the
desktop shortcut).
Program execution is a matter of following the on-screen instructions. The
program contains help screens to provide additional information you may need to
know to execute the steps. When the program asks whether you want to execute
the performance verification or the adjustment procedure, select the adjustment
procedure.

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals