Adjustment Procedures
5-8
TDS7104 & TDS7054 Service Manual
Figure 5- 3: National Instruments GPIB Interface Properties dialog box
h. Click on the device DEV1 in the Device Name list to select it.
i. Right click on the selection and select Rename.
j. Type in the new name DUT1.
k. Repeat the process for devices DEV2 through DEV7, renaming them
according the names in Table 5--4. These are the names that will be
recognized by the test program.
Table 5- 4: Renaming the GPIB devices
Initial name New n ame
DEV1 DUT1
DEV2 SG1
DEV3 FG1
DEV4 AUX1
DEV5 PS1
DEV6 PG1
DEV7 RT1
l. Click on the (new) device DUT1 in the Device Name dialog box.
m. Under DUT1 Attributes, set the Primary GPIB Address to 1.