Adjustment Procedures
5-6
TDS7104 & TDS7054 Service Manual
c. In the GPIB Configuration control window, click the Talk/Listen button
to select Talk/Listen mode.
d. Touch the Address readout and then click the up and down arrows or the
upper multipurpose knob to set the GPIB address to 1.
e. Click the Close button to close the control window.
4. If you are using a separate PC as the controller, do step 5 and its substeps
and then s kip step 6. If you are using the oscilloscope as the controller, skip
step 5 and then do step 6 and its substeps.
5. Configure the PC controller: Perform these steps if you are using a separate
PC as the controller. Use the GPIB software and hardware on the controller
as necessary to make the following settings:
a. Set the controller GPIB address to 0.
b. Define GPIB devices and their addresses for compatibility with the test
program as shown in Table 5--3.
Table 5- 3: GPIB devices required by test
program
Device name GPIB address
<controller> 0
DUT1 1
SG1 Set each address to that of
FG1
the F luke 9500B
AUX1
PS1
PG1
RT1
6. Configure the oscilloscope controller: Perform these steps if you are using
the oscilloscope as the controller.
NOTE. If you use the USB port, this procedure will take much longer.