Performance Tests
4-30
TDS7104 & TDS7054 Service Manual
TDS7000 Test Record
Photocopy this table and use it to record the performance test results for your
TDS7000 Series Oscilloscope.
TDS 7000 Test Record
Instrument Serial Number: Certificate Number:
Temperat ure: RH %:
Date of Calibration: Technician:
TDS7000 performance test Minimum Incoming Outgoing Maximum
DC vol tage measurement accuracy (averaged)
CH1 5 mV Vert scale setting,
--5 Div position setting, +1 V offset
+ 1. 0355 V __________ __________ + 1.0445 V
CH1 5 mV Vert scale setting,
+5 Div position setting, --1 V offset
-- 1.0445 V __________ __________ -- 1.0355 V
CH1 200 mV Vert scale setting,
--5 Div position setting, +10 V offset
+ 11.5195 V __________ __________ + 11.6805 V
CH1 200 mV Vert scale setting,
+5 Div position setting, --10 V offset
-- 11.6805 V __________ __________ -- 11.5195 V
CH1 1.01 V Vert scale setting,
--5 Div position setting, +10 V offset
+ 17. 621 V __________ __________ + 18.379 V
CH1 1.01 V Vert scale setting,
+5 Div position setting, --10 V offset
-- 18.379 V __________ __________ -- 17.621 V
CH2 5 mV Vert scale setting,
--5 Div position setting, +1 V offset
+ 1. 0355 V __________ __________ + 1.0445 V
CH2 5 mV Vert scale setting,
+5 Div position setting, --1 V offset
-- 1.0445 V __________ __________ -- 1.0355 V
CH2 200 mV Vert scale setting,
--5 Div position setting, +10 V offset
+ 11.5195 V __________ __________ + 11.6805 V
CH2 200 mV Vert scale setting,
+5 Div position setting, --10 V offset
-- 11.6805 V __________ __________ -- 11.5195 V
CH2 1.01 V Vert scale setting,
--5 Div position setting, +10 V offset
+ 17. 621 V __________ __________ + 18.379 V
CH2 1.01 V Vert scale setting,
+5 Div position setting, --10 V offset
-- 18.379 V __________ __________ -- 17.621 V
CH3 5 mV Vert scale setting,
--5 Div position setting, +1 V offset
+ 1. 0355 V __________ __________ + 1.0445 V
CH3 5 mV Vert scale setting,
+5 Div position setting, --1 V offset
-- 1.0445 V __________ __________ -- 1.0355 V
CH3 200 mV Vert scale setting,
--5 Div position setting, +10 V offset
+ 11.5195 V __________ __________ + 11.6805 V

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals