<10. Other Functions>
10-8
IM 11M12A01-04E 11th Edition : Jul. 19, 2017-00
10.3 Initialization Procedure
Follow the table below to initialize parameters. The password for initialization is 1255.
Table 10.5 Initialization Procedure
Switch operation Display Description
> ∧
ENT
F30
Enter the parameter code for the item to be initialized.
The following show an example of entering "F30."
(Previous needed operations are omitted.)
> ∧
ENT
0000
Touch the [ENT] key to switch to the password entry display.
> ∧
ENT
1000
Enter the password 1255 for initialization.
> ∧
ENT
1000
> ∧
ENT
1200
> ∧
ENT
1200
> ∧
ENT
1250
> ∧
ENT
1250
> ∧
ENT
1255
> ∧
ENT
1255
After you enter the password and then touch the [ENT] key, all the
digits ash.
> ∧
ENT
USr Go
Touch the [ENT] key again to display "USr Go."
> ∧
ENT
USr Go
Touch the [ENT] key once more. All the digits then ash for two to
three seconds, and data initialization starts.
> ∧
ENT
F30
The initialization is complete, and the parameter code selection
display then appears.
The symbol ( ) indicates that the keys are being touched, the light characters indicates ashing.
WARNING
Do not attempt to turn off the equipment power during initialization (while “USr Go” is ashing).
10.4 Reset
Resetting enables the equipment to restart. If the equipment is reset, the power is turned off
and then back on. In practical use, the power remains on, and the equipment is restarted under
program control. Resetting will be possible in the following conditions: