TECHNICAL REFERENCE MANUAL
5.7.2. HS general specifications
ADC
Type
16-bit SAR
Sampling Rate
Simultaneous 1 MS/sec
Analog anti alias filter
Bypass or analog 100 kHz 5th order Bessel
Digital low pass filter
Software selectable
Digital filter Characteristic
Bypass, Butterworth or Bessel
Filter Order
2nd , 4th, 6th or 8th
Ratio Sample rate to Filter Freq.
From 2 to 100
Topology
Cascaded IIR Filter (up to 4 sections)