Monitoring NBP
2000966-386D Dash⢠3000/4000/5000 10-9
Custom Defaults
Setup Custom Automatic NBP Measurements
To define up to four individual automatic NBP measurement series, follow this
procedure.
CAUTION
āChanges made to these control settings are persistent. They are
recalled when a patient is discharged from the monitor.
NOTE
Only available in software version 6 or later.
1. Select the NBP parameter window label.
2. Select Setup Custom Auto.The Setup Custom Auto window displays. There are
four BP SERIES sections
3. To define the first measurement series, follow this procedure.
a. Select 1ST BP SERIES.
b. Select the time interval between measurements from the available options (1
minute to 4 hours).
c. Select REPEAT.
d. Select the number of times to repeat the measurement (1to 25 times or
continuously until manually stopped).
4. Repeat 3 for each series (2ND, 3RD and 4TH).
5. To begin the automatic measurement cycle, select START CUSTOM AUTO. The
measurement cycle begins with the first series, completes each measurement the
defined number of times then proceeds to the second, third and fourth series. If
any series is not defined or is turned off, it is skipped and the next series in the
cycle is started.
6. To change a series while in progress, select the NBP parameter window label and
select SETUP SERIES.
ī To stop one of the defined series, select the SERIES > OFF.
ī To stop measurements for one of the defined series, select REPEAT > OFF.
Auto NBP Cancellation Notification
If the monitor is set to acquire measurements automatically (NBP AUTO), and a
time-out or no determination occurs, automatic measurement is cancelled, the
message āAUTO OFFā displays in the NBP parameter window and a system status
alarm will sound.