3-4 Dashâ„¢ 3000/4000/5000 2000966-386D
Monitor Setup
 COMBO — Acquire ECG data from a monitor or a telemetry receiver cabinet
and access all the available parameters from the monitor. A network connection
is required.
 ROVER COMBO — Combines the mobility of ROVER mode and the telemetry
compatibility of COMBO mode.
To determine the mode running on a monitor, select MORE MENUS > MONITOR
SETUP > SOFTWARE CONFIGURATION. The monitoring mode is listed under
ADMIT MENU TYPE.
Patient-Monitor Type
There are three patient-monitor types. The patient-monitor type controls the monitor
defaults and monitor settings. It is defined during installation via the password
protected SERVICE MODE menu.
îš„ ADULT-ICU
îš„ NEONATAL-ICU
îš„ OPERATING ROOM
To determine the patient-monitor type running on a monitor, select MORE MENUS >
MONITOR SETUP > SOFTWARE CONFIGURATION. The patient-monitor type
is listed under PATIENT MONITOR TYPE.
For more information on the monitor default settings for each patient-monitor type,
refer to Factory Defaults on page C-1.
Language
The monitor language is set to one of 14 options when installed. In most cases, the
language does not affect the monitor functionality or usability. However, in a few
languages, you may experience the following conditions:
NOTE
Special characters that are sent over the network in alarm broadcasts, patient
status or system status messages, patient, bed or care unit names cannot be
displayed by networked equipment that does not support these languages.
 HUNGARIAN, POLISH and RUSSIAN — Special characters may appear as
blanks, asterisks or square boxes due to the inability to display Cyrillic (Russian)
characters, as well as certain Polish and Hungarian characters.
 CHINESE and JAPANESE — All text for alarm broadcast sent over the
network is in English. All text input (for example, unit name, bed name, patient
information, etc.) is also in English.
 CHINESE — Invasive and non-invasive pressure parameter values are displayed
as mmHg. A converted kilopascals (kPa) value also displays in the invasive and
non-invasive pressure parameter windows when kPa is turned on. To set the
default display of the kPa values to ON or OFF, select the MONITOR
DEFAULTS > SETUP DEFAULT DISPLAY > NBP/IBP KPA UNITS setting.
To individually turn on or turn off the display of kPa units, select the KPA
UNITS control setting from the pressure parameter menu. Choosing OFF
displays the parameter value in mmHg.